Fisherbrand Lab Furnishings Catalog

Lab Design Guidelines

General Plumbing Details

Minimum Spacing Between Fixtures The drawing below shows the allowable minimum spacing between service fittings on laboratory bench top.

4 1/2” (11.4 cm)

7” (17.8 cm)

9” (22.9 cm)

10 1/2” (26.7 cm)

12” (30.5 cm)

7” (17.8 cm)

Service and Rough-In Designations and Color Coding

Recommended Supply Line Pressures

Designation

Service

Color

Service

Pressure

E

Electric (120 Volt AC)

-

Hot and Cold Water

Local Pressure

DT

Data

-

Gas

0.216 psi

A

Air

Orange

Air (Ordinary)

15 psi

G

Gas

Dark Blue

Air (Air Powered Stirrer Operation)

40 psi

V

Vacuum

Yellow

Vacuum (Ordinary)

15”-22” of Mercury

N

Nitrogen

Brown

Vacuum (High)

28.5” of Mercury

O

Oxygen

Light Green

Steam (Low Pressure)

0-15 psi as required

S

Steam

Black

Steam (High Pressure)

15-30 psi as required

CW

Cold Water

Green

HW

Hot Water

Red

DW Distilled Water

White

W

Waste

-

WV

Waste Vent

-

254

Made with FlippingBook - professional solution for displaying marketing and sales documents online