Fisherbrand Lab Furnishings Catalog
Lab Design Guidelines
General Plumbing Details
Minimum Spacing Between Fixtures The drawing below shows the allowable minimum spacing between service fittings on laboratory bench top.
4 1/2” (11.4 cm)
7” (17.8 cm)
9” (22.9 cm)
10 1/2” (26.7 cm)
12” (30.5 cm)
7” (17.8 cm)
Service and Rough-In Designations and Color Coding
Recommended Supply Line Pressures
Designation
Service
Color
Service
Pressure
E
Electric (120 Volt AC)
-
Hot and Cold Water
Local Pressure
DT
Data
-
Gas
0.216 psi
A
Air
Orange
Air (Ordinary)
15 psi
G
Gas
Dark Blue
Air (Air Powered Stirrer Operation)
40 psi
V
Vacuum
Yellow
Vacuum (Ordinary)
15”-22” of Mercury
N
Nitrogen
Brown
Vacuum (High)
28.5” of Mercury
O
Oxygen
Light Green
Steam (Low Pressure)
0-15 psi as required
S
Steam
Black
Steam (High Pressure)
15-30 psi as required
CW
Cold Water
Green
HW
Hot Water
Red
DW Distilled Water
White
W
Waste
-
WV
Waste Vent
-
254