Government Scientific Source Lab Furnishings

Fume Hoods

High Performance Fume Hood Technical Data: • Our hoods at UL1805 listed and are tested in accordance to ASHRAE-110. • All hoods are pre-wired to a junction box on the roof of the hood with two 120 volt AC 20 amp GFCI receptacles on the side posts, T8-fluorescent light fixture with bulb, 20 amp light switch and a duplex for an alarm on the roof. • Each front post and interior end liner is punched and capped for up to five (5) remote control service fittings. • The left hand post is punched and covered for a second electrical fixture at the top which may be used for a fan switch or other electrical device. The right post is punched and covered for an alarm. • Service fittings, fan switch, work top, cup sink, alarm, and base units must be ordered separately. •

Seismic restraints are available upon request. High Performance Bench Hood Roof Detail Center Line Center Line of Duct 10 1/4”

24”

48”

24”

Center Line of Duct 10 1/4”

Stainless Steel Duct Collar

Alarm Outlet Pre-wired Junction Box

Pre-wired Junction Box

T-8 Lights

T-8 Light

48”, 60” ,72”

96”

Left Section Detail

Hood and Duct Size Chart Hood Size

Stainless Steel Duct Collar

Duct Size

32 15/16

48" 60" 72" 96"

1:10" 1:12" 1:12" 2:10"

62 15/16” 57 1/8”

35 1/8” 48”

43 3/8”

24”

24” Clear Sash Opening

8”

22”

32 15/16”

39

703-734-1805