Cole-Parmer Lab Design & Build Program
Lab Design Guidelines
General Plumbing Details
Minimum Spacing Between Fixtures The drawing below show the allowable minimum spacing between service fittings on laboratory bench top.
4 1/2” (11.4 cm)
7” (17.8 cm)
9” (22.9 cm)
10 1/2” (26.7 cm)
12” (30.5 cm)
7” (17.8 cm)
Service and Rough-In Designations and Color Coding
Recommended Supply Line Pressures
Service
Pressure
Designation
Service
Color
Hot and Cold Water
Local Pressure
E
Electric (120 Volt AC)
- -
Gas
0.216 psi
DT
Data
Air (Ordinary)
15 psi
A G V N O S
Air
Orange
Air (Air Powered Stirrer Operation) Vacuum (Ordinary) Vacuum (High) Steam (Low Pressure) Steam (High Pressure)
40 psi
Gas
Dark Blue
18-22" of Mercury 28.5" of Mercury
Vacuum Nitrogen Oxygen
Yellow Brown
Light Green
0-15 psi as required
Steam
Black Green
15-30 psi as required
CW HW
Cold Water Hot Water
Red
DW Distilled Water
White
W
Waste
- -
WV
Waste Vent
www.coleparmer.com
90
Made with FlippingBook HTML5