Fisherbrand Quick Ship Catalog CA
Hoods and Enclosures
Chemical Fume Hoods
Standard Bench Fume Hood Technical Data
• Fisherbrand™ Hoods are UL1805 listed and are tested in accordance to ASHRAE-110. • All hoods are pre-wired to a junction box on the roof of the hood with two 120 volt AC 20 amp GFCI receptacles on the side posts, T8-fluorescent light fixture with bulb, 20 amp light switch and a duplex for an alarm on the roof • Each front post and interior end liner is punched and capped for up to five (5) remote control service fittings. • The left hand post is punched and covered for a second electrical fixture at the top which may be used for a fan switch or other electrical device. The right post is punched and covered for an alarm. • Service fittings, fan switch, work top, cup sink, alarm, and base units must be ordered separately.
• Seismic restraints are available upon request. Standard Bench Hood Roof Detail Center Line
Center Line of Duct 10 1/4”
Hood and Duct Size Chart
Stainless Steel Duct Collar
Alarm Outlet Pre-wired Junction Box
Hood Size Duct Size
48”
1:10”
T-8 Light
60”
1:12”
48”, 60” ,72”
72”
1:12”
Left Section Detail
35 7/8” 32 15/16”
3” Sash Extension Above Hood
Stainless Steel Duct Collar
35 1/8” 48” Clear Interior Rear Service Chase
57 1/8” Height of Hood
24” Hood Interior
43” Full View Sash Area 28” Clear Sash Opening
22”
8”
32 15/16” Depth of Hood with Upper Fascia Removed
34
Made with FlippingBook - professional solution for displaying marketing and sales documents online