Fisherbrand Lab Furnishings Catalog 04.29.22 FINAL

Hoods and Enclosures

Chemical Fume Hoods

High Performance Bench Fume Hood Technical Data

• Fisherbrand™ Hoods are UL1805 listed and are tested in accordance to ASHRAE-110. • All hoods are pre-wired to a junction box on the roof of the hood with two 120 volt AC 20 amp GFCI receptacles on the side posts, T8-fluorescent light fixture with bulb, 20 amp light switch and a duplex for an alarm on the roof • Each front post and interior end liner is punched and capped for up to five (5) remote control service fittings. • The left hand post is punched for a second electrical fixture at the top which may be used for a fan switch or other electrical device. The right post is punched and covered for an alarm. • Service fittings, fan switch, work top, cup sink, alarm, and base units must be ordered separately. • Seismic restraints are available upon request. High Performance Bench Hood Roof Detail Center Line Center Line of Duct 8 1/4” 26 3/8” 48” 26 3/8”

Center Line of Duct 8 1/4”

Stainless Steel Duct Transition

Alarm Outlet Pre-wired Junction Box

Pre-wired Junction Box

T-8 Lights

T-8 Light

48”, 60” ,72”

96”

Left Section Detail

Hood and Duct Size Chart

35 7/8” 32 15/16”

Hood Size Duct Size

Stainless Steel Round Duct Transition

48”

1:12”

60”

1:12”

72”

1:12”

96”

2:12”

35 1/8” 48” Clear Interior Rear Service Chase

57 1/8” Height of Hood

24” Hood Interior

43” Full View Sash Area 24” Clear Sash Opening 62 15/16” Height with Duct Transition

22”

8”

32 15/16” Depth of Hood with Upper Fascia Removed

86